The BIOHAZARD ATLANTIC CYTOCABINET 1200 hood is a microbiological safety cabinet suitable for the preparation of cytotoxic and chemotherapeutic drugs, and for the manipulation of biological agents, in hospital pharmacies, oncology departments, day hospitals and pharmaceutical industries.
It is a class II type H vertical laminar flow cabin, with front opening through which the operator can work inside the working chamber and which is designed and built to protect the operator, increase the protection of the product from external contamination, and to minimize biological risks for the environment. All this is made possible by the absolute triple filtration (hepa filters).
Accessories on request:
- Activated carbon cell to be cascaded after the Hepa filter for the expelled air for processing with toxic vapors.
- N.1 UV lamp to be placed inside the 30 Watt working chamber
- Chest of 3 drawers with wheels and key
- EXACARB module - Activated carbon filtration system.
Work area dimensions mm LxWxH
|
1130 x 550 x 570
|
Overall dimensions
mm LxWxH
|
1320 x 785 x 2270
|
Average speed vertical laminar flow (m / s)
|
0.40 |
Average speed front barrier flow (m / s)
|
0.45
|
Total / expelled air volume (mc / h.)
|
1350/400
|
Weight in Kg
|
270 |
Absorbed power W
(230V 50Hz)
|
700 + 440
|
Protection fuses
|
2 x 5 AF (5 x 20) mm
|
Watt / Lux lighting lamp |
1x36W 900 Lux |
Network connection socket |
10A |
UV germicidal lamp - accessory |
1 x 30W |
External dimensions of the product
Height: |
2270 |
mm |
Width: |
1320 |
mm |
Depth: |
785 |
mm |
Weight: |
270 |
Kg |